MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE (127 DEG) SIZE 2.5 ACCOLADE (127 DEG) SIZE 2.5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE (127 DEG) SIZE 2.5 ACCOLADE (127 DEG) SIZE 2.5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 6021-2530
Medical Device Problem Codes Detachment of Device or Device Component (2907); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Pain (1994); Ambulation or Postural Difficulties (2544); Metal Related Pathology (4530)
Date of Event 12/31/2021
Type of Reportable Event Serious Injury
Event or Problem Description
Right side hip revision, all components explanted were stryker components.The patient was re-implanted with new revision stryker components.The patient recently was reporting pain in her right hip, x-rays showed a disassociation and intraprosthetic dislocation of the femoral head from the trunnion of the stem.The patient¿s cobalt chrome levels were elevated and the stem/femoral head junction was failed, so the surgeon needed to revise the femoral component.There were adverse local tissue reactions surrounding the component, the femoral head had corrosion and severe wear was noted on the trunnion as the femoral head was fractured off.
 
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Device not returned.
 
Event or Problem Description
Right side hip revision, all components explanted were stryker components.The patient was re-implanted with new revision stryker components.The patient recently was reporting pain in her right hip, x-rays showed a disassociation and intraprosthetic dislocation of the femoral head from the trunnion of the stem.The patient¿s cobalt chrome levels were elevated and the stem/femoral head junction was failed, so the surgeon needed to revise the femoral component.There were adverse local tissue reactions surrounding the component, the femoral head had corrosion and severe wear was noted on the trunnion as the femoral head was fractured off.
 
Additional Manufacturer Narrative
Reported event: an event regarding disassociation and altr/abnormal ion levels/corrosion involving an accolade stem was reported.The disassociation event was confirmed, but the altr/abnormal ion levels/corrosion event was not confirmed.Method & results: -product evaluation and results: visual inspection: the reported device was not returned however photographs were provided for review.Visual inspection of the provided photographs indicated that the stem was recently explanted.The stem trunnion appears severely worn.Damage is consistent with loss of taper lock.-clinician review: a review of the provided medical information by a clinical consultant indicated: "conclusion of assessment: this inquiry reports the failure of a total hip replacement after over 15 years due to trunnion wear, elevated ion levels and disassociation.I can confirm that this event took place since i was able to see the explanted components.Regarding the possible root cause of this event, i cannot determine it with certainty.Causes of trunnion wear and disassociation are multifactorial including surgical technique factors, patient factors and implant factors." -product history review: review of the device history records indicate the devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to disassociation of the head from the stem, altr, elevated ion levels, and corrosion at the head-stem junction.The disassociation event was confirmed through medical review and visual inspection of provided photographs of the devices.The altr/abnormal ion levels/corrosion event was not confirmed.The exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the devices, pathology reports, the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE (127 DEG) SIZE 2.5 ACCOLADE (127 DEG) SIZE 2.5
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
anna ryan
raheen business park
limerick NA
EI   NA
61498200
MDR Report Key13314420
Report Number0002249697-2022-00119
Device Sequence Number17333621
Product Code LPH
UDI-Device Identifier04546540510747
UDI-Public04546540510747
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K121308
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2006
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2010
Device Model Number6021-2530
Device Catalogue Number6021-2530
Device Lot Number13671303
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/31/2021
Supplement Date Received by Manufacturer04/04/2022
Initial Report FDA Received Date01/21/2022
Supplement Report FDA Received Date04/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/16/2005
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age83 YR
Patient SexFemale
-
-