DEPUY ORTHOPAEDICS INC US PFC*SIGMA DI AUG16MMCO,SZ3LEFT; SIGMA REVISION IMPLANT : KNEE FEMORAL ACCESSORY Back to Search Results
Model Number 96-0871
Medical Device Problem Code Tear, Rip or Hole in Device Packaging (2385)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/15/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on (b)(6) 2021, during the resterilization and repackaging process, the operator identified multiple parts displayed a visual cosmetic defect on the top surface of the pfc sigma augments.The augments had foam particles transferred from the packaging components.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2021, during the resterilization and repackaging process, the operator identified multiple parts displayed a visual cosmetic defect on the top surface of the pfc sigma augments.The augments had foam particles transferred from the packaging components.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
B.Implants are available for return.C.The correct lot number is e37ae4.I believe these parts have been returned and received in warsaw.I have replied to this several times.I am not sure why it is not registering.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: examination of the returned sigma augment confirmed inner foam packaging residue/debris was found transferred to the surface of the device.The debris was found in raynham during repacking/resterilization of expired sterile product returned from the field.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the product and lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
SIGMA REVISION IMPLANT : KNEE FEMORAL ACCESSORY
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS 1818910
700 orthopaedic dr
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key13315042
Report Number1818910-2022-01375
Device Sequence Number11149918
Product Code JWH
UDI-Device Identifier10603295234340
UDI-Public10603295234340
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 01/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number96-0871
Device Catalogue Number960871
Device Lot Number182913
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/10/2022
Supplement Date Received by Manufacturer01/26/2022
03/31/2022
05/23/2022
Initial Report FDA Received Date01/21/2022
Supplement Report FDA Received Date02/07/2022
04/27/2022
05/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/11/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-