Model Number 482804645
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Pain (1994)
Date of Event 09/23/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It is reported that revision surgery was performed to remove an es2 screw that had migrated and was now causing the patient pain.During the revision surgery, while the surgeon was removing a different screw, the screw appeared to have fractured.Upon review of the pre-operative x-ray, the surgeon noticed that the other es2 screw had fractured before the revision surgery.Both screws were removed.This report captures the migrated screw.
 
Additional Manufacturer Narrative
Visual, dimensional, functional and material analysis could not be performed as the device was not returned.Device history records could not be reviewed as a valid lot number was not provided and could not be obtained.Complaint history records were reviewed for this catalog and no adverse trends were observed.The date of initial surgery is unknown.Revision surgery was performed on (b)(6) 2021.Patient's detail including activity level and bone quality are unknown.Operative notes and x-rays were not provided.Due to lack of information and device not returned, an exact cause of the migration of screw cannot be determined.Potential causes include: inadequate initial construct/ excess torque applied on the construct, poor bone quality of patient, patient non-fusion, high activity level, patient fall, etc.Status and location of the device is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ES2 INTEGRATED BLADE SCREW SIZE L 6.5X45MM
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
STRYKER SPINE-US
2 pearl court
allendale NJ 07401
Manufacturer (Section G)
STRYKER SPINE-FRANCE
zone industrielle de marticot
cestas 33610
FR   33610
Manufacturer Contact
rita karan
2 pearl court
allendale, NJ 07401
2017608000
MDR Report Key13315323
Report Number0009617544-2022-00005
Device Sequence Number1701408
Product Code NKB
UDI-Device Identifier07613327002188
UDI-Public07613327002188
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K122845
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number482804645
Device Catalogue Number482804645
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/23/2021
Initial Report FDA Received Date01/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-