| Model Number |
71026-70 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/07/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A customer reported receiving erroneous glucose results from an abbott diabetes care device.The results when plotted on a parkes error grid fell into either the c, d, or e zone.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Dhrs for the freestyle lite meter were reviewed and the dhrs showed the freestyle lite meter passed all tests prior to release.Dhrs for the freestyle strips were reviewed and the dhrs showed the freestyle strips passed all tests prior to release.Retain testing was performed for freestyle strips and all units performed in specification and passed.The reported meter has been returned and investigated.Visual inspection has been performed on the returned meter and no issues were observed.Meter powered on with button and test strips.Control solution testing has been performed and all results were within range specification.No malfunction or product deficiency was identified.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Search Alerts/Recalls
|