MICRO THERAPEUTICS, INC. DBA EV3 AXIUM PRIME BRPL 3D; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number APB-2-2-3D-ES
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/17/2022
Type of Reportable Event Malfunction
Event or Problem Description
Medtronic received a report that the axium prime coil could not be detached after 3 attempts with the instant detacher.Another instant detacher was not used, and manual detachment was not attempted.There were no problems with the instant detacher as it was successfully used with other coils before and after the non-detachment. the devices were prepared according to the instructions for use (ifu). the event was said to be not associated with the patient.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional information received reported vessel tortuosity was normal.Prior to coil non-detachment there were no issues encountered.No pushwire damage was observed after removal from the patient.
 
Additional Manufacturer Narrative
H3: as found condition- the axium prime coil device was returned for analysis within a shipping box, within three resealable plastic bi ohazard pouches.The opened axium prime inner pouch was also returned.The instant detacher used in the event was not returned.Visual inspection/damage location details: the actuator interface was found loose on the coupler tube.This is indicative of a detachment attempt using an instant detacher at this location.The break indicator was found intact.There is no evidence of manual detachment at this location.No damages or irregularities were found with the rest of the pusher.The coin was found still against the lumen stop.Blood was found under the ptfe shrink tubing and within the lumen stop.No damages or irregularities were found with the shield coil.The implant coil was found still attached to the pusher.The implant coil was found damaged.Testing/analysis ¿ the axium prime implant coil failed to detach by retracting the loose proximal segment or by breaking the break indicator and retracting the release wire as the release wire was found stuck.Under magnification, the ptfe shrink tubing was removed, and the blood was dissolved.The release wire was retracted, and the coin exited the lumen stop with no issues and no resistance encountered.No other damages or anomalies were found with the returned device.Conclusion - based on the customer report and device analysis, the customer report of "non-detachment" was confirmed.It is likely the cause of the non-detachment is the blood found within the lumen stop and under the ptfe shrink tubing causing the coin to become stuck within the lumen stop.The implant coil was found damaged.It is possible the implant became damaged when attempting to advance against resistance or damage during return shipping to medtronic as the device was returned outside its introducer sheath.There is no indication that the event is related to a potential manufacturing issue, therefore a device history record review was not required.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIUM PRIME BRPL 3D
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key13315832
Report Number2029214-2022-00078
Device Sequence Number11943728
Product Code KRD
UDI-Device Identifier00847536021174
UDI-Public00847536021174
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K151447
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/27/2023
Device Model NumberAPB-2-2-3D-ES
Device Catalogue NumberAPB-2-2-3D-ES
Device Lot Number221255964
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/20/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/17/2022
Supplement Date Received by Manufacturer02/03/2022
03/07/2022
Initial Report FDA Received Date01/21/2022
Supplement Report FDA Received Date03/01/2022
03/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/27/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-