| Model Number |
300866 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/28/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that bd¿ eccentric luer slip tip syringe ce mark sterile had a damaged tip.The following information was provided by the initial reporter: "broken barrel".
|
| |
|
Additional Manufacturer Narrative
|
|
A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
The following fields were updated due to additional information: d.9.Device available for eval?: yes.D.9.Returned to manufacturer on: 1/20/2022.H.6.Investigation: sample and photo received for investigation.Upon visual inspection, the tip appears damaged and deformed.A device history review was performed for lot 2105010 no deviations or non-conformances were identified during the manufacturing process that could have contributed to the reported issue.Possible root cause is associated with transport of the barrel from molding to assembly machine.
|
| |
|
Event or Problem Description
|
|
It was reported that bd¿ eccentric luer slip tip syringe ce mark sterile had a damaged tip.The following information was provided by the initial reporter: "broken barrel.".
|
| |
|
Search Alerts/Recalls
|