MAUDE Adverse Event Report: BECTON DICKINSON, S.A. BD¿ ECCENTRIC LUER SLIP TIP SYRINGE CE MARK STERILE,; PISTON SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON, S.A. BD¿ ECCENTRIC LUER SLIP TIP SYRINGE CE MARK STERILE,; PISTON SYRINGE Back to Search Results
Model Number 300866
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/28/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that bd¿ eccentric luer slip tip syringe ce mark sterile had a damaged tip.The following information was provided by the initial reporter: "broken barrel".
 
Additional Manufacturer Narrative
A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d.9.Device available for eval?: yes.D.9.Returned to manufacturer on: 1/20/2022.H.6.Investigation: sample and photo received for investigation.Upon visual inspection, the tip appears damaged and deformed.A device history review was performed for lot 2105010 no deviations or non-conformances were identified during the manufacturing process that could have contributed to the reported issue.Possible root cause is associated with transport of the barrel from molding to assembly machine.
 
Event or Problem Description
It was reported that bd¿ eccentric luer slip tip syringe ce mark sterile had a damaged tip.The following information was provided by the initial reporter: "broken barrel.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD¿ ECCENTRIC LUER SLIP TIP SYRINGE CE MARK STERILE,
Common Device Name
PISTON SYRINGE
Manufacturer (Section D)
BECTON DICKINSON, S.A.
camino de valdeolivia
s/n
san agustin de guadalix
Manufacturer (Section G)
BECTON DICKINSON, S.A.
camino de valdeolivia
s/n
san agustin de guadalix
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13319658
Report Number3003152976-2022-00010
Device Sequence Number1648393
Product Code FMF
UDI-Device Identifier30382903008668
UDI-Public30382903008668
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K980987
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date04/30/2026
Device Model Number300866
Device Catalogue Number300866
Device Lot Number2105010
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/28/2021
Supplement Date Received by Manufacturer02/25/2022
Initial Report FDA Received Date01/21/2022
Supplement Report FDA Received Date03/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-