Manufacturer¿s ref.No: (b)(4).Information regarding patient identifier, date of birth, age, gender, weight, race, and ethnicity were not provided.Procode is krd/hcg.The initial reporter phone: (b)(6).The initial reporter email address is not available / reported.[conclusion]: the healthcare professional reported during a coil embolization procedure with the target located on the lumbar artery, the 4mm x 15cm deltafill 18 (dlf180415 / 30536998) was the first coil used.The coil was energized for detachment, but it was not energized.The physician made several attempts but could not detach the coil.The physician replaced the concomitant enpower control cable (ecb00018200 / 30666648) and attempted to detach the coil again, but the same issue occurred.Subsequently, the coil was replaced with another coil from another lot number) and the procedure was continued.The coil was energized and successfully detached.The procedure was resumed to completion.There was no adverse event nor patient injury associated with the issue reported.It was indicated that the product is not available to be returned for evaluation and analysis.Additional information was received on 16-jan-2022.The information indicated that the coil was successfully removed from the patient.The coil was still attached to the delivery system when it was removed.A pre-deployment electrical check was performed.All connections appeared to fit properly without application of excessive force.The information indicated that the procedure was completed with another 4mm x 15cm deltafill 18 coil of a different lot number.Based on complaint information, the device is not available to be returned for analysis.A review of manufacturing documentation associated with this lot (30536998) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no nonconformances related to device manufacture or inspection.All product rejected during manufacturing was identified as scrap and properly accounted for.Product analysis cannot be conducted as the product was not returned for analysis.No determination of causes and possible contributing factors could be made.As such, the investigation will be closed.With the information available and without the product available for analysis, the reported customer complaint could not be confirmed.Based on the manufacturing documentation review, there is no indication that the event is related to the device manufacturing process.Assignment of root cause for the event remains speculative and inconclusive, based on the limited information provided and without the return of the complaint sample; however, it is possible that clinical and procedural factors, including device manipulation / interaction, aneurysm size / vessel characteristics, device selection, and the concomitant microcatheter, may have contributed to the reported issue.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective / preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
|
|
The healthcare professional reported during a coil embolization procedure with the target located on the lumbar artery, the 4mm x 15cm deltafill 18 (dlf180415 / 30536998) was the first coil used.The coil was energized for detachment, but it was not energized.The physician made several attempts but could not detach the coil.The physician replaced the concomitant enpower control cable (ecb00018200 / 30666648) and attempted to detach the coil again, but the same issue occurred.Subsequently, the coil was replaced with another coil from another lot number) and the procedure was continued.The coil was energized and successfully detached.The procedure was resumed to completion.There was no adverse event nor patient injury associated with the issue reported.It was indicated that the product is not available to be returned for evaluation and analysis.Additional information was received on (b)(6) 2022.The information indicated that the coil was successfully removed from the patient.The coil was still attached to the delivery system when it was removed.A pre-deployment electrical check was performed.All connections appeared to fit properly without application of excessive force.The information indicated that the procedure was completed with another 4mm x 15cm deltafill 18 coil of a different lot number.
|