| Model Number |
93024-132 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The reported hemofilter subject of the event is not available to be returned for evaluation.The reported product is indicated for single use and was exchanged upon detection of the filter leak.No issue was noted with the replacement hemofilter.A review of manufacturing records indicates that the reported filter met all pre-release specifications regarding sterility, bacterial endotoxin testing, biocompatibility, product integrity and dimensions.The reported event did not cause or contribute to any serious injury or deterioration of health and is not expected to lead to serious injury if the event were to recur.Medivators will continue to monitor for similar events to ensure the product continues to perform as expected.
|
| |
|
Event or Problem Description
|
|
The user facility reported that during a procedure involving the medivators hf1200 hemofilter, a filter leak occurred.The reported filter leak resulted in an approximately 100ml volume of patient blood lost.The filter was exchanged and the procedure was completed successfully without delays.No additional reports of malfunction or injury were reported.
|
| |
|
Search Alerts/Recalls
|