Model Number 2200-2010
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/04/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The patient's dob, weight, ethnicity, and race are unknown.This information was not available from the facility.Patient information regarding relevant tests/laboratory data is unknown.This information was not available from the facility.Foreign: (b)(6).Visual inspection found no unusual characteristics of the device.During functional testing, using an indeflator filled with green color dye water, the device was pressurized and at less than 2 atm, a pinhole leak was present in the middle of the balloon.The ifu warns at least 99.9% of the balloons (with a 95% confidence level) will not burst at or below the rbp.
 
Event or Problem Description
The angiosculpt device was used to treat a severely calcified mid lad.During the second inflation at 18 atm, the balloon ruptured.The procedure was completed with a non-angiosculpt device.No patient injury reported.This product problem is being submitted conservatively because the balloon ruptured below rbp.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIOSCULPT RX PTCA
Common Device Name
CATHETER, PTCA, CUTTING/SCORING
Manufacturer (Section D)
SPECTRANETICS
5055 brandin court
fremont CA 94538
Manufacturer (Section G)
SPECTRANETICS
5055 brandin court
fremont CA 94538
Manufacturer Contact
ana tan
5055 brandin court
fremont, CA 94538
MDR Report Key13323225
Report Number3005462046-2022-00004
Device Sequence Number12017799
Product Code NWX
UDI-Device Identifier00813132021146
UDI-Public00813132021146
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
P050018
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/14/2023
Device Model Number2200-2010
Device Catalogue Number2200-2010
Device Lot NumberG20010017
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/10/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/04/2022
Initial Report FDA Received Date01/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/14/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ASAHI: 0.014" GUIDEWIRE; MEDTRONIC: 6F GUIDE CATHETER
Patient Age64 YR
Patient SexMale
-
-