Model Number 6290
Medical Device Problem Code Power Problem (3010)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/02/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the communicator power cord was burning and melted.The patient advocate called that about a month ago or so woke up and smelled something burning.The communicator power plug melted and prongs into the receptacle.The patient advocate had to replace the socket and needed pliers to get it out.The company representative opted to replace the power cord.The communicator remains in service while the power cord will be returned for analysis.No adverse patient effects were reported.
 
Additional Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event or Problem Description
It was reported that the communicator power cord was burning and melted.The patient advocate called that about a month ago or so woke up and smelled something burning.The communicator power plug melted and prongs into the receptacle.The patient advocate had to replace the socket and needed pliers to get it out.The company representative opted to replace the power cord.The communicator remains in service while the power cord will be returned for analysis.No adverse patient effects were reported.
 
Additional Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, additional information from product investigation indicates that the returned power adapter is no longer functioning.Device had been subject to excess heat induced by the patient's environment, likely a faulty power source.The brick plate that holds the prongs of power adapter had begun to melt and bubble and could no longer hold prongs in place for use.As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LATITUDE COMMUNICATOR
Common Device Name
EXTERNAL COMMUNICATOR
Manufacturer (Section D)
JABIL CIRCUIT (SHANGHAI) LTD
600 tian lin rd
shanghai
CH 
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key13327509
Report Number2124215-2021-38028
Device Sequence Number13865152
Product Code LWS
UDI-Device Identifier00802526544224
UDI-Public00802526544224
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P910077/S140
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number6290
Device Catalogue Number6290
Device Lot Number5190136
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/02/2021
Supplement Date Received by Manufacturer04/06/2022
Initial Report FDA Received Date01/22/2022
Supplement Report FDA Received Date05/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient Age78 YR
Patient SexMale
-
-