H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing-related cause for this event.Investigation summary: one crosser catheter loaded onto a sidekick support catheter was returned for evaluation.During the visual evaluation, it was noted that the distal tip was misaligned and separated from the outer catheter.Therefore, the investigation is confirmed for material separation.A definitive root cause for the reported material separation issue could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiration date: 10/2023), g3, h6(method).H11: h6(result, conclusion).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|