• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS SHENZHEN MAGNETIC RESONANCE LTD. ARTIS ONE; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS SHENZHEN MAGNETIC RESONANCE LTD. ARTIS ONE; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Model Number 10848600
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/17/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
 
Event or Problem Description
It was reported to siemens that an adverse event occurred while operating the artis one system.At the beginning of a procedure, a cover from the display ceiling suspension system fell and hit the patient.The procedure was continued and completed.Following the procedure, the patient was observed by the nursing team and did not present any complications.After observation, the patient was released on the same day.Siemens has requested additional information in order to conduct an investigation of the reported event.
 
Additional Manufacturer Narrative
Siemens has completed an investigation of the reported event.The root cause was determined to be improper workmanship.Visual inspection showed that two buckles of the cover peeled off which led to unsatisfactory fixation of the cover assembly.The cover had been disassembled/reassembled several times in the past year for general service activities.It is assumed that the two missing buckles may have been broken during service activities but were undiscovered at that time.The cover assembly was replaced during service activities.Furthermore, the maintenance instructions will be optimized to detail the steps to further guide the service/installation personnel to ensure all the buckles of the cover are in good status before reassembling the cover.After detailed investigation, the incident is not classified as a reportable event as neither serious injury, death nor an unexpected, prolonged hospitalization of the patient or any other person occurred or could be expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTIS ONE
Common Device Name
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
SIEMENS SHENZHEN MAGNETIC RESONANCE LTD.
gaoxin c ave.
2nd, hi-tech industrial park
shenzhen, china 51805 7
CH  518057
Manufacturer (Section G)
SIEMENS SHENZHEN MAGNETIC RESONANCE LTD.
gaoxin c ave.
2nd, hi-tech industrial park
shenzhen, china 51805 7
CH   518057
Manufacturer Contact
meredith adams
40 liberty blvd.
malvern, PA 19355
6104486461
MDR Report Key13332638
Report Number3004754211-2022-12982
Device Sequence Number20054929
Product Code OWB
UDI-Device Identifier04056869009964
UDI-Public04056869009964
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K133580
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Remedial Action Inspection,Repair
Type of Report Initial,Followup
Report Date (Section B) 01/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number10848600
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/17/2022
Supplement Date Received by Manufacturer03/10/2022
Initial Report FDA Received Date01/24/2022
Supplement Report FDA Received Date03/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-