Model Number TFNT00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Visual Disturbances (2140); Halo (2227)
Date of Event 12/01/2021
Type of Reportable Event Serious Injury
Event or Problem Description
A facility representative reported that following an intraocular lens (iol) implant procedure, the patient experienced poor vision and halos.The iol was exchanged for another lens model four nd half months following the initial implant procedure.The patient was not hospitalized for the event and there was no patient harm.The prognosis was good.No more information available.
 
Additional Manufacturer Narrative
The lens was returned for evaluation.Solution is dried on the lens.The lens was cleaned for further evaluation.A scratch is observed on the anterior surface of the optic.Product history records were reviewed and the documentation indicated the product met release criteria.Qualified associated products were indicated.The root cause for the reported events could not be determined.A malfunction has not been indicated or information provided that the lens was the cause of the event.The file indicates the 23.5 diopter (add power +2.2/+3.2) lens was replaced with a toric (1.50cyl), 22.5 diopter (1.50 extended depth of focus) lens.Due to the exchange of a non-toric lens to a toric lens, this may suggest that the replacement lens model was an improved choice for the patient's vision needs.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ PANOPTIX TRIFOCAL IOL
Common Device Name
LENS, MULTIFOCAL
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key13333065
Report Number1119421-2022-00110
Device Sequence Number20055080
Product Code MFK
UDI-Device Identifier00380652415064
UDI-Public00380652415064
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
PMA/510(K) Number
P040020
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Reporter Type Manufacturer
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 01/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/09/2025
Device Model NumberTFNT00
Device Catalogue NumberTFNT00U235
Device Lot Number15019402
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/12/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/28/2021
Initial Report FDA Received Date01/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D; MONARCH III IOL DELIVERY SYSTEM, INJECTOR; PROVISC OPHTHALMIC VISCOSURGICAL DEVICE
Outcome Attributed to Adverse Event Required Intervention;
Patient Age59 YR
Patient SexFemale
-
-