| Model Number |
1570-01-100 |
| Medical Device Problem Codes |
Osseointegration Problem (3003); Migration (4003)
|
| Health Effect - Clinical Codes |
Fall (1848); Bone Fracture(s) (1870); Inadequate Osseointegration (2646)
|
| Date of Event |
01/11/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient had a fractured hip that was fixed with a summit stem and a bipolar head in (b)(6).The patient suffered a fall and broke her greater trochanter that caused her stem to subside.The doctor decided to remove the current implants and replaced them with a 8¿ solution stem and a new bipolar head.He also cabled the trochanter to help reduce the fracture.There was loosening of the femoral component (non cemented) noted as well.No surgical delay.Doi: (b)(6) 2021.Dor: (b)(6) 2022 (left hip).
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|