STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON P/A CR BEADED #6L; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Model Number 5517-F-601
Medical Device Problem Code Osseointegration Problem (3003)
Health Effect - Clinical Codes Pain (1994); Inadequate Osseointegration (2646)
Date of Event 12/28/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate that devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Device not returned.
 
Event or Problem Description
It was reported that the patient's left knee was revised due to pain and femoral loosening.Surgeon reported that a possible cause or contributor is that the original cuts made with the mako robot seemed off.A cr femur and 7x9 cs insert were revised to a ts femur with stem and 7x9 ps insert.Rep reported that pictures can be provided, and confirmed that no further information will be released by the hospital or surgeon.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON P/A CR BEADED #6L
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NA
EI   NA
Manufacturer Contact
alessandra chavez
2555 davie road
fort lauderdale, FL 33317
9546280700
MDR Report Key13333172
Report Number0002249697-2022-00124
Device Sequence Number19231364
Product Code MBH
UDI-Device Identifier07613327040975
UDI-Public07613327040975
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2026
Device Model Number5517-F-601
Device Catalogue Number5517F601
Device Lot NumberL627J
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/28/2021
Initial Report FDA Received Date01/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age61 YR
Patient SexMale
Patient Weight147 KG
-
-