ARTHREX, INC. BUR,OVAL,CLEARCUT,12 FLUT,5.5MMX 13CM; BURR, SURGICAL, GENERAL & PLASTIC SURGERY
|
Back to Search Results |
|
| Model Number |
BUR,OVAL,CLEARCUT,12 FLUT,5.5MMX 13CM |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/04/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Complaint confirmed.Even though the device was not returned for evaluation, picture provided by the customer showed the clear tip hood has melted.The cause of this event is undetermined; however, most likely caused by prying and/or leveraging the device for a prolonged use without proper irrigation.
|
| |
|
Event or Problem Description
|
|
It was reported that during a surgery the sleeve around the tip of the device melted.No part of the device broke off.According to the surgeon no harm for patient, operator or third party occurred.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint confirmed.Even thought the device was not returned for evaluation, picture provided by the customer showed the clear tip hood has melted.The cause of this event is undetermined; however, most likely caused by prying and/or leveraging the device for a prolonged use without proper irrigation.
|
| |
|
Search Alerts/Recalls
|
|
|