MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET 360 NANO 1MM X 3CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET 360 NANO 1MM X 3CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Model Number M0035421030
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/27/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the procedure, the subject coil encountered strong resistance at the hub of the microcatheter.It was replaced with another coil, but resistance was also noted with this coil.The coil was replaced the third time but it still could not be advanced inside the catheter.Upon removal, the physician confirmed that the subject coil was prematurely deployed inside the microcatheter and was stretched at the proximal section.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.It was reported that during the procedure, the subject coil encountered strong resistance at the hub of the microcatheter.Upon removal, it was confirmed that the subject coil was prematurely deployed inside the microcatheter and was stretched at the proximal section.It was replaced with another coil, but resistance was also noted with this coil.The coil was replaced for the third time, but still it could not be advanced inside the catheter.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Additional Manufacturer Narrative
The device is not available to the manufacturer.
 
Additional Manufacturer Narrative
Due to the automated mes system there are controls in the manufacturing process to ensure the product met specifications upon release.Visual and functional inspection were not performed because the product was not returned for investigation.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.Additional information provided by the customer indicated that there was friction or resistance while the coil was advancing the introducer sheath.The friction or resistance were felt at the hub.The devices were prepared for use as per the directions for use.The devices were confirmed to be in good condition during preparation/prior to use on the patient.Continuous flush was set up and maintained throughout the clinical procedure.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause and the device was not returned, an assignable cause of undeterminable will be assigned to this complaint.
 
Event or Problem Description
It was reported that during the procedure, the subject coil encountered strong resistance at the hub of the microcatheter.It was replaced with another coil, but resistance was also noted with this coil.The coil was replaced the third time but it still could not be advanced inside the catheter.Upon removal, the physician confirmed that the subject coil was prematurely deployed inside the microcatheter and was stretched at the proximal section.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.It was reported that during the procedure, the subject coil encountered strong resistance at the hub of the microcatheter.Upon removal, it was confirmed that the subject coil was prematurely deployed inside the microcatheter and was stretched at the proximal section.It was replaced with another coil, but resistance was also noted with this coil.The coil was replaced for the third time, but still it could not be advanced inside the catheter.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET 360 NANO 1MM X 3CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key13333684
Report Number3008881809-2022-00033
Device Sequence Number1595011
Product Code HCG
UDI-Device Identifier04546540697974
UDI-Public04546540697974
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/06/2023
Device Model NumberM0035421030
Device Catalogue NumberM0035421030
Device Lot Number22492894
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/27/2021
Supplement Date Received by Manufacturer03/23/2022
Initial Report FDA Received Date01/24/2022
Supplement Report FDA Received Date04/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/07/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
EXCELSIOR SL-10 MICROCATHETER (STRYKER).; INZONE DETACHMENT SYSTEM (STRYKER).; TARGET COILS (STRYKER).
Patient Age74 YR
Patient SexFemale
-
-