• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LDR MÉDICAL ROI-C IMPLANT 14X14 H7; ROI-C TITANIUM-COATED IMPLANT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LDR MÉDICAL ROI-C IMPLANT 14X14 H7; ROI-C TITANIUM-COATED IMPLANT SYSTEM Back to Search Results
Catalog Number MC1312P
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/24/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that an roi-c construct was installed too deep, then the plates and cage were damaged when being backed out of the disc space.There were no patient impacts.This is report two of two for this event.
 
Additional Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference report 3004788213-2022-00006.
 
Additional Manufacturer Narrative
Corrections in d4: udi number, d9, and h3.Additional information in d4: expiration date, h4 and h6: component, investigation type, findings, and conclusions.Device evaluation: visual inspection revealed that the cage was fractured and the anchoring plates also have material deformation on the bottom side likely related to removal.Root cause: root cause was unable to be determined.It is possible that the space was not fully distracted, or the patient had hard bone causing difficulty placing the cage in the right place.Dhr review: per dhr reviews, the parts were likely conforming when they left zimvie control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
 
Event or Problem Description
It was reported that an roi-c construct was installed too deep, then the plates and cage were damaged when being backed out of the disc space.There were no patient impacts.This is report two of two for this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROI-C IMPLANT 14X14 H7
Common Device Name
ROI-C TITANIUM-COATED IMPLANT SYSTEM
Manufacturer (Section D)
LDR MÉDICAL
quartier europe de l¿ouest
5, rue de berlin
sainte-savine, na 10300
FR  10300
Manufacturer (Section G)
LDR MÉDICAL
quartier europe de l¿ouest
5, rue de berlin
sainte-savine, na 10300
FR   10300
Manufacturer Contact
kim martinez
10225 westmoor dr.
na
westminster, CO 80021
3035144809
MDR Report Key13333711
Report Number3004788213-2022-00007
Device Sequence Number1595599
Product Code OVE
UDI-Device Identifier03662663016070
UDI-Public(01)03662663016070(17)240901(10)77876
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K151934
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/01/2024
Device Catalogue NumberMC1312P
Device Lot Number77876
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/27/2021
Supplement Date Received by Manufacturer06/17/2022
Initial Report FDA Received Date01/24/2022
Supplement Report FDA Received Date06/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/22/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
Patient Weight79 KG
-
-