BECTON DICKINSON MEDICAL (SINGAPORE) BD LUER-LOK¿ SYRINGE WITH BD ECLIPSE¿ DETACHABLE NEEDLE; HYPODERMIC SINGLE LUMEN NEEDLE Back to Search Results
Model Number 305785
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/03/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
There were multiple pma / 510(k)#s reported to be involved.The information for each 510(k) number is as follows: pma / 510(k)#: k980987 (syringe).Pma / 510(k)#: k161170 (needle).A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the plunger in the bd luer-lok¿ syringe with bd eclipse¿ detachable needle was broken.The following information was provided by the initial reporter: "broken plunger".
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval: yes.D10: returned to manufacturer on: 2022-01-13.H6: investigation summary: one photo and one sample were received by our quality team for evaluation.From the photo and sample, the plunger was observed to be cracked at the diameter c area.A review of the internal manufacturing device records and raw material history files for the reported lot number was performed and no recorded quality problems or rejections to this incident were found.The manufacturing process was reviewed.The probable root cause could be due to the plunger falling from a high height from the chute bin when transferring from the molding to the assembly area.An extra chute bin was installed to cushion part drop before the good part bin after the reported batch was produced.This incident has been added to our database of reported incidents.
 
Event or Problem Description
It was reported that the plunger in the bd luer-lok¿ syringe with bd eclipse¿ detachable needle was broken.The following information was provided by the initial reporter: "broken plunger.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD LUER-LOK¿ SYRINGE WITH BD ECLIPSE¿ DETACHABLE NEEDLE
Common Device Name
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
Manufacturer (Section G)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13334037
Report Number8041187-2022-00029
Device Sequence Number13864629
Product Code FMI
UDI-Device Identifier30382903057857
UDI-Public30382903057857
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
SEE H10
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date03/31/2026
Device Model Number305785
Device Catalogue Number305785
Device Lot Number1118018
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/13/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/03/2022
Supplement Date Received by Manufacturer02/14/2022
Initial Report FDA Received Date01/24/2022
Supplement Report FDA Received Date03/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-