MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION V-18 CONTROL WIRE; WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION V-18 CONTROL WIRE; WIRE, GUIDE, CATHETER Back to Search Results
Model Number 40535
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 11/24/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that guidewire tip detachment occurred.Physician was performing a lower extremity pulmonary vein isolation (pvi).A 300cm, 12cm v-18 control wire was used to cross the lesion.However, during the procedure, the tip of the guidewire was fractured.The device was then removed but the tip was retained in the vessel.The detached tip could not be retrieved.No patient complications were reported after the tip detachment and the patient fully recovered.
 
Additional Manufacturer Narrative
Unit was returned with its original pouch batch 27836205.The returned device matches with upn provided by the customer.The returned guidewire was kinked at the distal tip section.Additionally, the core wire was exposed due to peeling of the polymer jacket and was detached at the distal tip.The peeled section related to the polymer was measured and 1.2cm was missing.Microscopic inspection revealed the distal tip was bent and the polymer peeled and detached.The very tip is without detachment or an exposed tip.The device was measured in three sections, and was within specification.
 
Event or Problem Description
It was reported that guidewire tip detachment occurred.Physician was performing a lower extremity pulmonary vein isolation (pvi).A 300cm, 12cm v-18 control wire was used to cross the lesion.However, during the procedure, the tip of the guidewire was fractured.The device was then removed but the tip was retained in the vessel.The detached tip could not be retrieved.No patient complications were reported after the tip detachment and the patient fully recovered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V-18 CONTROL WIRE
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13334089
Report Number2134265-2022-00252
Device Sequence Number19244531
Product Code DQX
UDI-Device Identifier08714729070238
UDI-Public08714729070238
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K033742
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/13/2023
Device Model Number40535
Device Catalogue Number40535
Device Lot Number0027836205
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/17/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/30/2021
Supplement Date Received by Manufacturer01/25/2022
Initial Report FDA Received Date01/24/2022
Supplement Report FDA Received Date02/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-