• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE ICONIX 2 NEEDLES WITH INTELLIBRAID TECHNOLOGY, 2.3MM ANCHOR WITH 2.0MM XBRAID TT; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE ICONIX 2 NEEDLES WITH INTELLIBRAID TECHNOLOGY, 2.3MM ANCHOR WITH 2.0MM XBRAID TT; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number 3910500222
Health Effect - Clinical Codes Device Embedded In Tissue or Plaque (3165); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/27/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the tip of the inserter broke during the procedure and remains imbedded in the bone.
 
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Additional Manufacturer Narrative
Alleged failure: the tip of the inserter broke off and was buried in the bone and could not be removed.The failure(s) identified in the investigation is consistent with the complaint record.The probable root causes could be: use of excessive force, movement of guide out of orientation to pilot hole, or use of wrong drill.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
 
Event or Problem Description
It was reported that the tip of the inserter broke during the procedure and remains imbedded in the bone.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICONIX 2 NEEDLES WITH INTELLIBRAID TECHNOLOGY, 2.3MM ANCHOR WITH 2.0MM XBRAID TT
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
andrea zenere
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key13334975
Report Number0002936485-2022-00048
Device Sequence Number17159832
Product Code MBI
UDI-Device Identifier37613327283465
UDI-Public37613327283465
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K151201
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number3910500222
Device Lot Number21222AE2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/12/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/27/2021
Supplement Date Received by Manufacturer12/27/2021
Initial Report FDA Received Date01/24/2022
Supplement Report FDA Received Date03/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/10/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-