Evaluation of the returned device found the fio2 readings are out of specifications and the mixer balance was out of calibration.Upon further inspection of the device, it was noted the set screws have be replaced with non-sechrist parts.The fio2 knob shaft has multiple marks on it indicating the knob was reset and the face of the mixer shows signs of the stop pin making contact with the face of the unit.A dhr review found no evidence that a manufacturing non-conformity contributed to the reported complaint, and the instructions for use were reviewed and determined to provide adequate instructions and warnings for the safe and effective use of the device.Therefore, no corrective or preventative actions are necessary.All complaints are trended and reviewed by management on a monthly basis.As part of this monthly review, any excursion above the control limits for this failure mode will be assessed, documented and acted upon as warranted.Manufacturer reference file no.(b)(4).
|