Model Number 196
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Codes Low Oxygen Saturation (2477); Respiratory Insufficiency (4462)
Date of Event 12/24/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The actual complaint product was not returned for evaluation.A review of the device history record is in-progress.All information reasonably known as of 19-jan-2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).
 
Event or Problem Description
Avanos medical, inc.Received a single report that referenced two different incidences, which were associated with separate units, involving two different events.This is the first of two reports.Refer to 8030647-2022-00004 for the second event.Immediately after installing the device saline began going into my patient's lungs instead of the suction system.The patient began to decompensate and the oxygen level desaturated.The device was replaced and the patient was quickly suctioned.
 
Additional Manufacturer Narrative
The device history record for the reported lot number, 30144156, in this complaint was reviewed and the material was produced according to the manufacturing procedures and met the quality requirements.The device was evaluated and the failure was not confirmed.Root cause could not be determined.All information reasonably known as of 17-mar-2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21 cfr 803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
 
Additional Manufacturer Narrative
Correction: lot # 30144156.The product involved in the report has been returned and is being processed for evaluation.A review of the device history record is in-progress.All information reasonably known as of 10-feb-2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BALLARD CS NEO/PED, 6 FR, Y-ADAPTER
Common Device Name
VAP CLOSED SUCTION CATHETERS & ACCESSORIES
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT S DE RL DE CV
carretera internacional
salida norte no. 1053
magdalena, cp
Manufacturer Contact
lisa clark
5405 windward parkway
alpharetta, GA 30004
4704485444
MDR Report Key13335048
Report Number8030647-2022-00003
Device Sequence Number5023391
Product Code BSY
UDI-Device Identifier00609038938240
UDI-Public00609038938240
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/08/2026
Device Model Number196
Device Catalogue NumberN/A
Device Lot Number30144156
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/27/2021
Supplement Date Received by Manufacturer01/18/2022
02/25/2022
Initial Report FDA Received Date01/24/2022
Supplement Report FDA Received Date02/10/2022
03/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age16 DA
Patient SexMale
-
-