• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INSTITUT STRAUMANN AG S-IMPLANT Ø 4.8MM WN, SLACTIVE 6MM; ENDOSSEOUS DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INSTITUT STRAUMANN AG S-IMPLANT Ø 4.8MM WN, SLACTIVE 6MM; ENDOSSEOUS DENTAL IMPLANT Back to Search Results
Model Number SLActive Titanium TL
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 10/16/2021
Type of Reportable Event Serious Injury
Event or Problem Description
The clinician reports the implant was inserted (b)(6) 2009 in fdi 46.On (b)(6) 2021, loss of osseointegration was verified.The device was forwarded to the manufacturer.At the event the patient experienced: infection and peri-implantitis.No further patient complications were reported.
 
Additional Manufacturer Narrative
The batch number could be verified.Our manufacturing q-system assures, that production and process controls are in place to ensure that batches confirm to the applicable specifications before they are distributed.  the loss of an endosseous dental implant after successful osseointegration and restoration is a known inherent risk of the procedure.Implants may have to be removed in case one or more of the implant success criteria are not met.Implant success criteria according to buser et al.(1991) are: absence of persistent subjective complaints such as pain, foreign body sensation and /or dysesthesia.Absence of a recurrent peri-implant infection with suppuration.Absence of implant mobility.Absence of a continuous radiolucency around the implant.The manufacturer's trend analysis confirms that the reported failure rate associated with its dental implants is below the expected failure rate for this treatment as published in the scientific literature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S-IMPLANT Ø 4.8MM WN, SLACTIVE 6MM
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
INSTITUT STRAUMANN AG
peter merian-weg 12
basel 04002
SZ  04002
Manufacturer (Section G)
INSTITUT STRAUMANN AG
peter merian-weg 12
basel 04002
SZ   04002
Manufacturer Contact
jennifer jackson
60 minuteman road
andover, MA 01810
9787472509
MDR Report Key13335966
Report Number0009613348-2022-09287
Device Sequence Number4885394
Product Code DZE
UDI-Device Identifier07630031716323
UDI-Public07630031716323
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K033984
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 01/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2012
Device Model NumberSLActive Titanium TL
Device Catalogue Number033.630S
Device Lot NumberG3945
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/30/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date01/24/2022
Event Location Other
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date01/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/30/2007
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
Patient SexFemale
-
-