U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL ARTHSCO,4/140_30_QIK/STRKR HUB; RIGID ENDOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL ARTHSCO,4/140_30_QIK/STRKR HUB; RIGID ENDOSCOPE Back to Search Results
Model Number 242043
Medical Device Problem Codes Break (1069); Poor Quality Image (1408); Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/19/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported by the sales rep that preoperatively to an unknown surgery on (b)(6) 2021, it was observed that the distal tip on the arthsco,4/140_30_qik/strkr hub endoscope device came into contact with the shaver.During in-house engineering evaluation, it was determined that there were minor scratches, poor quality and dark image, distal tip damaged, burnt, and shaved down and window and optics were damaged and fiber recessed on the device.Another like device was used to complete the procedure.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Additional Manufacturer Narrative
Additional narrative: udi: (01)10886705026234(21)25127 investigation summary: the complaint device was received at the service center and evaluated.It was reported that the distal tip came into contact with shaver.Per service reports, this complaint can be confirmed.During the service evaluation the following defects were identified: minor scratches on the unit.Functional : poor image quality.Distal tip damaged, burnt, and shaved down.Window and optics damaged.Image dark.Fiber recessed.The defective parts were replaced to resolve the issues.After repair, the device was found to be working according to the specifications.The faulty parts was identified as the root cause for the device failure during the service evaluation.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device history lot: manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTHSCO,4/140_30_QIK/STRKR HUB
Common Device Name
RIGID ENDOSCOPE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ  CH-2400
Manufacturer Contact
kara ditty-bovard
chemin-blanc 38
le locle CH-24-00
SZ   CH-2400
6013142063
MDR Report Key13336585
Report Number1221934-2022-00270
Device Sequence Number11741936
Product Code HRX
UDI-Device Identifier10886705026234
UDI-Public10886705026234
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K971996
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number242043
Device Catalogue Number242043
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/04/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/19/2022
Initial Report FDA Received Date01/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-