BOSTON SCIENTIFIC CORPORATION EMERGE; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7133
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/20/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The 85% stenosed, 2mmx28mm was located in the moderately tortuous and moderately calcified left anterior descending artery.A 2.00mm x 15mm emerge balloon catheter was advanced for use.However, during the procedure, the balloon ruptured.The procedure was completed with another of the same device.No patient complications nor injuries reported and the patient status was stable.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of an emerge mr balloon catheter.The device was visually and microscopically examined.At 52.5cm and 72.7cm from the strain relief, the hypotube of the device was kinked.The balloon was loosely folded.The device was attached to an inflation device filled with water and was inflated to rated burst pressure.The device inflated and deflated with no issue or irregularity.Product analysis could not confirm reported event as the device inflated to rated burst pressure and deflated with no issue.
 
Event or Problem Description
It was reported that balloon rupture occurred.The 85% stenosed, 2mmx28mm was located in the moderately tortuous and moderately calcified left anterior descending artery.A 2.00mm x 15mm emerge balloon catheter was advanced for use.However, during the procedure, the balloon ruptured.The procedure was completed with another of the same device.No patient complications nor injuries reported and the patient status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMERGE
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13336738
Report Number2134265-2022-00578
Device Sequence Number3919099
Product Code LOX
UDI-Device Identifier08714729806196
UDI-Public08714729806196
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K113220
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/26/2024
Device Model Number7133
Device Catalogue Number7133
Device Lot Number0027755679
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/11/2022
Supplement Date Received by Manufacturer03/23/2022
Initial Report FDA Received Date01/24/2022
Supplement Report FDA Received Date03/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age67 YR
Patient SexFemale
Patient Weight69 KG
-
-