U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LAKE REGION MEDICAL ZIPWIRE HYDROPHILIC GUIDEWIRE; WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LAKE REGION MEDICAL ZIPWIRE HYDROPHILIC GUIDEWIRE; WIRE, GUIDE, CATHETER Back to Search Results
Model Number M00146316B1
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/01/2021
Type of Reportable Event Malfunction
Event or Problem Description
Event description: per email, it was reported that: injury type: other during the procedure, it was found that the tip of the guide wire was peeling off, immediately replaced with a new guide wire, the patient continued treatment.Complaint summary: it was reported that the guidewire tip separation occurred.A guidewire was selected for use.However, during the procedure, it was found that the tip of the guide wire was peeling off.The procedure was completed with a different device.There were no patient complications nor injuries reported.Additional information received january 18, 2022: what was the name of the procedure(s) performed? only knew the procedure is for angiography.Was it the first time the device was used? yes.Was there visible damage to the device and/or its packaging prior to use? yes any resistance felt during insertion? no.Was the catheter stuck/locked on the guidewire? no or was there just significant resistance between the devices (not stuck)? no.Was the guidewire hydrated or rinsed with saline solution prior to removal from the dispenser? yes.Was there an issue with lubricity of the device? yes.Did the physician encounter any resistance while advancing the zipwire to its intended location? no.Did the physician encounter any resistance while removing the zipwire? yes.Was the zipwire advanced through a metal cannula or needle? none.Was the zipwire flushed during the procedure? yes did any part of the guidewire detach inside the patient? no.Is the metal core wire exposed? no is there an image available? no.
 
Additional Manufacturer Narrative
The device was not received for evaluation at the time of this report; therefore, no physical analysis of the device can be performed.A review of the device history records of the specimen device does not present any indication of manufacturing defect or anomaly that could have impacted on the event as reported.During manufacturing the production operators are instructed to 100% visually and tactilely inspect for any obvious defect, which includes a tactile examination of the entire length of each wire.In addition, during packaging of this product the operators are instructed to 100% visually inspect for any obvious defect prior to shipment.The history records indicate this product was final inspection tested at lake region medical and was determined to be acceptable.As indicated in the device instructions for use, warnings: to prevent possible tissue damage, care should be taken when manipulating a device over a zipwire hydrophilic guide wire during the device's placement or withdrawal.If resistance is felt during device placement, discontinue the procedure and determine the cause of resistance before proceeding.If the cause of resistance cannot be determined, remove the zipwire hydrophilic guide wire and device as a unit to prevent possible damage and/or complications.If any resistance is felt or if the tip's behavior and/or location seems improper, stop manipulating the zipwire¿ hydrophilic guide wire and/or catheter and determine the cause under fluoroscopy.Failure to exercise proper caution may result in bending, kinking, separation of the guide wire's tip, damage to the catheter or damage to the vessel." the complaint is non-verifiable as the product was not returned for evaluation.At this time, it is not possible to assign a definitive root cause for the event as reported.Based on the information provided, it appears that clinical and/or procedural factors impacted on the event as reported.If there is any further relevant information provided, a follow up medwatch report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIPWIRE HYDROPHILIC GUIDEWIRE
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
LAKE REGION MEDICAL
340 lake hazeltine drive
chaska MN 55318
Manufacturer (Section G)
VENUSA DE MEXICO S. DE R.L. DE C.V.
calle hertz 1525
parque industrial a j bermudez
ciudad juarez, chihuahua 32470
MX   32470
Manufacturer Contact
victoria foss
340 lake hazeltine drive
chaska, MN 55318
9526418506
MDR Report Key13337096
Report Number9680001-2022-00002
Device Sequence Number11787324
Product Code DQX
UDI-Device Identifier08714729754862
UDI-Public08714729754862
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K000011
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/12/2024
Device Model NumberM00146316B1
Device Catalogue Number41656
Device Lot NumberJRZ6280983
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/05/2022
Initial Report FDA Received Date01/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-