• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON, S.A. 50 ML BD¿ ECCENTRIC LUER SLIP TIP SYRINGE CE MARK STERILE; PISTON SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON, S.A. 50 ML BD¿ ECCENTRIC LUER SLIP TIP SYRINGE CE MARK STERILE; PISTON SYRINGE Back to Search Results
Model Number 300866
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/04/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported while using 50 ml bd¿ eccentric luer slip tip syringe ce mark sterile the stopper was found to be deformed.The following information was provided by the initial reporter: distorted stopper.
 
Additional Manufacturer Narrative
A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Additional Manufacturer Narrative
A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Event or Problem Description
It was reported while using 50 ml bd¿ eccentric luer slip tip syringe ce mark sterile the stopper was found to be deformed.The following information was provided by the initial reporter: distorted stopper.
 
Event or Problem Description
It was reported while using 50 ml bd¿ eccentric luer slip tip syringe ce mark sterile the stopper was found to be deformed.The following information was provided by the initial reporter: distorted stopper.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval yes.D10: returned to manufacturer on: 2022-01-17.H6: investigation summary sample and photos received for investigation.Upon visual inspection, the stopper assembled incorrectly is observed.No further damage can be seen.A device history review was performed for reported lot 2105010, no deviations or non-conformances were identified during the manufacturing process that could have contributed to this reported issue.Possible root cause is associated when there is a misalignment of plunger-stopper-barrel in the assembly station during assembly process.In assembly stations of 50ls manufacturing lines there is a vision system to detect stoppers bad assembled or missing stopper.In case any is defective it is rejected automatically to scrap.In this case, vision system has not detected the failure, and no rejection was made.Final products in this manufacturing line, for this reference are sampled and they are subjected to visual and functional inspections during the different manufacturing sub-processes according to procedures.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
50 ML BD¿ ECCENTRIC LUER SLIP TIP SYRINGE CE MARK STERILE
Common Device Name
PISTON SYRINGE
Manufacturer (Section D)
BECTON DICKINSON, S.A.
camino de valdeolivia
s/n
san agustin de guadalix
Manufacturer (Section G)
BECTON DICKINSON, S.A.
camino de valdeolivia
s/n
san agustin de guadalix
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13337830
Report Number3003152976-2022-00017
Device Sequence Number11778311
Product Code FMF
UDI-Device Identifier30382903008668
UDI-Public30382903008668
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K980987
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date04/30/2026
Device Model Number300866
Device Catalogue Number300866
Device Lot Number2105010
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/17/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/04/2022
Supplement Date Received by Manufacturer02/16/2022
Initial Report FDA Received Date01/24/2022
Supplement Report FDA Received Date03/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-