Model Number 398493
Medical Device Problem Code Failure to Interrogate (1332)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/08/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
We received your event description for the above mentioned device and would like to thank you for supporting our post-market surveillance.As of today, the medical device is not available for analysis, therefore the device itself could not be investigated.The information you provided has been entered into our quality system as a complaint.These types of complaints are used to evaluate systems and device performance throughout our organization and help to maintain and improve the performance of our devices.Should additional relevant information or the device itself become available, the investigation will be updated.
 
Event or Problem Description
Device explanted due to failure to interrogate and device not transmitting to home monitoring.No adverse patient events reported.Should additional information become available, it will be added to this file.
 
Additional Manufacturer Narrative
The device was received for analysis.Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this device were re-investigated and all production steps were performed accordingly.In particular, the final acceptance test proved the device functions to be as specified.The device could not be interrogated with a programmer.Therefore the device was opened and subjected to further investigation.The visual inspection of the inner assembly showed no anomalies.The measurement of the battery voltage confirmed a depleted battery.In a next step, the battery was disconnected from the electronic module.Further thorough investigation of the electronic module did not show any anomalies.In particular the current consumption proved to be normal and expected.The battery was sent to the manufacturer for further analysis.The manufacturing records of the battery were inspected, documenting that the battery parameters were within specification during the manufacturing.The battery was subjected to a visual and an electrical inspection as well as a microcalorimetry analysis.The visual inspection of the battery did not reveal any external signs of damage.The measurement of the battery voltage confirmed a depleted battery and the microcalorimetry analysis showed a normal heat dissipation.In a next step, the battery was opened for destructive analysis.During analysis no internal electrical short was evident.However, the occurrence of a temporary short circuit that contributed to an increased internal self-depletion cannot be excluded.In summary, the root cause for the premature battery depletion could not be determined conclusively.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOMONITOR 2-AF
Common Device Name
IMPLANTABLE CARDIAC MONITOR
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key13338000
Report Number1028232-2022-00418
Device Sequence Number11736783
Product Code MXD
UDI-Device Identifier04035479135522
UDI-Public04035479135522
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K152995
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2022
Reporter Type Manufacturer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2020
Device Model Number398493
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/21/2022
Supplement Date Received by Manufacturer05/16/2022
Initial Report FDA Received Date01/24/2022
Supplement Report FDA Received Date05/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/07/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age70 YR
Patient SexUnknown
-
-