The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: material rupture, seroma.Manufacturer¿s reference number: pc-(b)(4).
|
|
It was reported that a (b)(6) year old african american female patient, who's undergone primary breast reconstruction with two 750cc artoura breast tissue expanders, suffered spontaneous left breast implant deflation and left breast asymmetry and seroma, post-procedure.The complications were diagnosed during an in-office visit.As a result, the patient underwent removal and replacement with a 750cc artoura breast tissue expander (catalog: smxp150rh lot: 9413923 sn: (b)(4) on (b)(6) 2021.
|
|
On february 14, 2022, the product investigation was completed.Device investigation summary: the product was returned to mentor for evaluation.Mentor then conducted a visual inspection, microscopic examination, and thickness measurement of the returned device.Visual analysis of the returned sample revealed that the artoura ss, t exp, hp,750cc tissue expander was found to be ruptured.Microscopic examination was performed and the cause of the deflation could not be identified.Therefore a thickness measurement was conducted on the shell at the tear, and the thickness was within manufacturing specifications.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Although no conclusion could be reach on the cause of the reported event, the instructions for use contain the following caution: causes of deflation of tissue expander include but are not limited to the following events: crease fold failure, intraoperative or postoperative trauma, excessive stresses or manipulations as may occur during normal, daily routines including customary and purposeful trauma such as that which can occur during vigorous exercise, athletics, etc.A seroma is a build-up of fluid around the device.Symptoms from a seroma may include swelling, pain, and bruising.A seroma may occur soon after surgery, or any time later on, which would be referred to as a late seroma.While the body may absorb seromas, some may require surgery for drainage.The american society of plastic surgeons recommends and encourages member surgeons to always submit breast devices, capsule, and effusion to pathology for examination.As part of mentor¿s quality process all devices are manufactured, inspected, and released to approved specifications.
|