• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN 840 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN 840 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 4-840120DIUU-US
Medical Device Problem Code No Flow (2991)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/06/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic has not received the suspect device/component from the customer for evaluation nor has the device been evaluated by the service engineer.The hospital biomedical engineer sourced a part and replaced the oxygen proportional solenoid valve assembly (psol) and reported that they had repaired the ventilator in-house.A review of the device history record identified that the device had no failures which are relevant to the failure reported by the field.All finished product testing requirements were deemed acceptable.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that while in use on a patient, this 840 ventilator generated a background check breath delivery unit (bdu) diagnostic message indicating a loss of ventilation.The patient was removed from the ventilator and placed on an alternate ventilator without injury.
 
Additional Manufacturer Narrative
H3: it was reported that while in use on a patient, this pb840 ventilator generated a background check breath delivery unit (bdu) diagnostic message indicating a loss of ventilation.The customer's biomedical engineer checked and found relevant diagnostic code and replaced an oxygen psol (proportional solenoid).Additionally, customer replaced the battery to resolve an unrelated diagnostic code.The ventilator passed all functional tests and calibrations.No product was returned for evaluation or made available to medtronic.The likely cause of the event was isolated to a potential fault in oxygen psol.A review of the device history record identified that the device had no failures which are relevant to the failure reported by the field.All finished product testing requirements were deemed acceptable.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it has met all medtronic quality "specifications." medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
840 VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
COVIDIEN
micheal collins rd mervue
galway
EI 
Manufacturer (Section G)
COVIDIEN
micheal collins rd mervue
galway
EI  
Manufacturer Contact
kelly adams
2101 faraday avenue
carlsbad, CA 92008
7066035046
MDR Report Key13338224
Report Number8020893-2022-00015
Device Sequence Number19858779
Product Code CBK
UDI-Device Identifier10884522019354
UDI-Public10884522019354
Combination Product (Y/N)N
Initial Reporter StateNH
Initial Reporter CountryUS
PMA/510(K) Number
K970460
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 07/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number4-840120DIUU-US
Device Catalogue Number4-840120DIUU-US
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/04/2022
Supplement Date Received by Manufacturer07/07/2022
Initial Report FDA Received Date01/24/2022
Supplement Report FDA Received Date07/08/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/27/2005
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-