Model Number 346367
Medical Device Problem Code Impedance Problem (2950)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/16/2021
Type of Reportable Event Malfunction
Event or Problem Description
Lead explanted due to an increase in impedance measurement of this lead.No adverse patient events reported.Should additional information become available, it will be added to this file.
 
Additional Manufacturer Narrative
The device is currently not available for analysis.No conclusion can be drawn at this time.No additional information is available at the moment.The file is closed.The investigation will be re-opened should additional data become available.
 
Additional Manufacturer Narrative
Upon receipt, the lead under complaint was found cut 48 cm proximal to the lead tip.The distal fragment was received for analysis.The proximal fragment including the is-1 connector pin was not returned.An extraction aid was found on the distal fragment, it is reasonable to assume that the lead was cut during the surgery.The analysis showed that the insulation was found abraded through between 30 cm and 35 cm proximal to the lead tip.The insulation damage in this area was consistent with a crushing type of movement.Due to the location and type of damage, analysis determined it was likely caused by entrapment in the clavicle first-rib region.Furthermore, the lead tip was found damaged, which most likely resulted from the extraction procedure due to tensile stress.Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this device were re-investigated.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.In conclusion, the analysis did not reveal any sign of a material or manufacturing problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELOX ST 60
Common Device Name
PACING LEAD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key13338443
Report Number1028232-2022-00422
Device Sequence Number4164684
Product Code NVN
UDI-Device Identifier04035479077853
UDI-Public04035479077853
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P950037
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2014
Device Model Number346367
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/24/2022
Supplement Date Received by Manufacturer01/28/2022
04/25/2023
Initial Report FDA Received Date01/24/2022
Supplement Report FDA Received Date01/31/2022
04/27/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/23/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age61 YR
Patient SexMale
-
-