Model Number RBY2C0512-A
Medical Device Problem Codes Physical Resistance/Sticking (4012); Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the renal artery using ruby coils, a non-penumbra microcatheter and a non-penumbra catheter.During the procedure, the physician placed a ruby coil in the target vessel using the microcatheter.While advancing the next ruby coil into the microcatheter, the physician experienced resistance but, the physician continued to advance the ruby coil further into the microcatheter.Subsequently, the physician continued to experience resistance.While moving the pusher assembly of the ruby coil, the physician noticed that the ruby coil did not respond.Subsequently, it was noticed that the ruby coil had unintentionally detached.Therefore, the physician successfully flushed the detached ruby coil into the target vessel using a syringe.The procedure was completed using two additional ruby coils and the same microcatheter.There was no report of an adverse effect to this patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBY COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key13339099
Report Number3005168196-2022-00023
Device Sequence Number19852175
Product Code HCG
UDI-Device Identifier00814548012896
UDI-Public00814548012896
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/24/2022
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberRBY2C0512-A
Device Catalogue NumberRBY2C0512
Device Lot NumberF88944
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/29/2021
Initial Report FDA Received Date01/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/12/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age78 YR
Patient SexMale
Date Report Sent to FDA01/01/2005
-
-