MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. SENSAR IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. SENSAR IOL; INTRAOCULAR LENS Back to Search Results
Model Number AR40M
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/17/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that an intraocular lens (iol) had bent haptic.No other information was provided.
 
Additional Manufacturer Narrative
Age, weight, ethnicity: unknown/ not provided.If implanted, give date: not applicable, as there is no information to indicate that the lens was implanted.If explanted, give date: not applicable, as there is no information to indicate that the lens was implanted; therefore not explanted.The device is not received; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Attempts have been made to obtain missing information; however, no definitive response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Section d9.Device available for evaluation? yes.Returned to manufacturer on: jan.14, 2022.Section h3.Device evaluated manufacturer? yes.Device evaluation: visual inspection under magnification revealed viscoelastic residue on the optic body and haptics.The lens was cleaned and, haptic damage was observed.The complaint issue could be confirmed; however, it may be attributed to handling, suggested by the presence of viscoelastic residue, and cannot be confirmed to be related to a manufacturing issue.Therefore, no product deficiency could be identified.Manufacturing record review: the manufacturing process record was evaluated and revealed that the product was manufactured and released according to specifications.A search revealed that no other complaints were received from this production order.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSAR IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key13339431
Report Number3012236936-2022-00195
Device Sequence Number19874945
Product Code HQL
UDI-Device Identifier05050474514232
UDI-Public(01)05050474514232(17)230319
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/19/2023
Device Model NumberAR40M
Device Catalogue NumberAR40MN0015
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/14/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/04/2022
Supplement Date Received by Manufacturer03/01/2022
Initial Report FDA Received Date01/24/2022
Supplement Report FDA Received Date03/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/19/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-