Age, weight, ethnicity: unknown/ not provided.If implanted, give date: not applicable, as there is no information to indicate that the lens was implanted.If explanted, give date: not applicable, as there is no information to indicate that the lens was implanted; therefore not explanted.The device is not received; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Attempts have been made to obtain missing information; however, no definitive response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
|
Section d9.Device available for evaluation? yes.Returned to manufacturer on: jan.14, 2022.Section h3.Device evaluated manufacturer? yes.Device evaluation: visual inspection under magnification revealed viscoelastic residue on the optic body and haptics.The lens was cleaned and, haptic damage was observed.The complaint issue could be confirmed; however, it may be attributed to handling, suggested by the presence of viscoelastic residue, and cannot be confirmed to be related to a manufacturing issue.Therefore, no product deficiency could be identified.Manufacturing record review: the manufacturing process record was evaluated and revealed that the product was manufactured and released according to specifications.A search revealed that no other complaints were received from this production order.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|