• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE (127 DEG) SIZE 5.5 ACCOLADE (127 DEG) SIZE 5.5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE (127 DEG) SIZE 5.5 ACCOLADE (127 DEG) SIZE 5.5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 6021-5537
Medical Device Problem Codes Material Erosion (1214); Detachment of Device or Device Component (2907); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Metal Related Pathology (4530)
Date of Event 06/15/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported through the filing of a lawsuit that allegedly the patient was implanted with an lfit anatomic cocr v40 femoral head on his right hip on or about (b)(6) 2011 and was revised on (b)(6) 2021.It is further alleged that he suffered injuries as a result of implantation and explantation of the device(s) at issue, device recall and excessive levels of chromium and cobalt in his blood.
 
Additional Manufacturer Narrative
The information in this report was provided by stryker orthopaedics legal affairs department.No additional information is available at this time due to the ongoing litigation.Should additional information become available, the evaluation summary will be submitted in a supplemental report.
 
Additional Manufacturer Narrative
An event regarding disassociation, fretting/metallosis, and abnormal ion level involving a accolade stem that was mated with an lfit v40 cocr head was reported.The disassociation event was confirmed by clinician review of provided medical records.The fretting/metallosis and abnormal ion level events were not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: a review of the provided medical information by a clinical consultant indicated: "i have reviewed the documentation related to pi including the product inquiry summary, an implant sheet from the index surgery dated (b)(6) 2011 and an operative report from the revision surgery dated 6/15/2021.[.] the revision surgery operative report confirms that this procedure was completed and that the reason for revision surgery was dissociation of the femoral head from the trunnion.The surgeon noted metallosis throughout the joint and marked wear of the femoral neck.No information was provided to document excessive serum metallic ions.The root cause for this event can not be determined from this limited information." product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the reported accolade stem was mated with an lfit v40 cocr head.The cocr head has been identified to be within scope of nc and capa.No further investigation for this event is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
It was reported through the filing of a lawsuit that allegedly the patient was implanted with an lfit anatomic cocr v40 femoral head on his right hip on or about (b)(6) 2011 and was revised on june 15, 2021.It is further alleged that he suffered injuries as a result of implantation and explantation of the device(s) at issue, device recall and excessive levels of chromium and cobalt in his blood.Update: review of the provided medical records by a clinician indicated the following: "the reason for revision surgery was dissociation of the femoral head from the trunnion.The surgeon noted metallosis throughout the joint and marked wear of the femoral neck.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE (127 DEG) SIZE 5.5 ACCOLADE (127 DEG) SIZE 5.5
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
joann ripoli
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13340753
Report Number0002249697-2022-00126
Device Sequence Number11741967
Product Code LPH
UDI-Device Identifier04546540510778
UDI-Public04546540510778
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K121308
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/31/2016
Device Model Number6021-5537
Device Catalogue Number6021-5537
Device Lot Number35364903
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/04/2022
Supplement Date Received by Manufacturer04/14/2022
Initial Report FDA Received Date01/25/2022
Supplement Report FDA Received Date05/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/12/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age67 YR
Patient SexMale
-
-