• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL CADD; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL CADD; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 21-7040-24
Medical Device Problem Code Air/Gas in Device (4062)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluation: the device was returned for investigation in used conditions, without its original packaging.Leak test: sample was tested using a syringe with colored water.Results: during the test, leak was found between luer and tube, only one sample.The complaint is confirmed.The root cause could was determined to be linked to manufacturing process.Root cause is lack of solvent by not following the established procedure.Awareness notification was made to production personnel in order to explain the importance to adherence or following in the procedure (b)(4) "extension set with filter assembly" section 7.0 bond male luer to coil assembly, by quality engineer on (b)(6) 2021.
 
Event or Problem Description
It was reported the device displayed alarm -air in line-.While priming medicine, it does not make it through the extension set and generates high pressure alarm.Extension set was changed and infusion started correctly.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
ofc 518
minneapolis, MN 55442
MDR Report Key13341653
Report Number3012307300-2022-01552
Device Sequence Number3918794
Product Code FPA
UDI-Device Identifier10610586020223
UDI-Public10610586020223
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K942046
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date11/19/2024
Device Model Number21-7040-24
Device Catalogue Number21-7040-24
Device Lot Number3910168
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/21/2021
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/14/2021
Initial Report FDA Received Date01/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/05/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-