| Model Number |
314-13-22 |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
12/22/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by manufacturer? pending evaluation.Concomitant medical device(s): 300-62-02 - stemless humeral comp integrip, cage, size 2; 310-61-44 - stemless humeral head 44mm x 17mm x alpha.
|
| |
|
Event or Problem Description
|
|
As reported, approximately 3 years post op the initial left tsa, this (b)(6) female patient's was revised due to aseptic glenoid loosening.Patient had a subscap failure and a loose tsa glenoid in stemless implant.The case report form indicates this event is possibly related to devices and/or procedure.This event report was received through clinical data collection activities.
|
| |
|
Additional Manufacturer Narrative
|
|
Section h10: (d4) serial number: (b)(6), expiration date: 24-jun-2023 (h3) the revision reported was likely the result of both patient-related conditions and an insufficient bond between the cage glenoid and the bone, which led to aseptic (non-infected) glenoid loosening.However, this cannot be confirmed as the device was not available for evaluation and pre-revision x-rays were not provided.(h4) device manufacture date: 26-jun-2018.Section h11: *the following sections have corrected information: (d10) concomitant device(s): 300-62-02, 5483855 - stemless humeral comp integrip, cage, size 2, 310-61-44, 5351181 - stemless humeral head 44mm x 17mm x alpha.
|
| |
|