BECTON DICKINSON, S.A. BD ECLIPSE¿ SAFETY NEEDLE; HYPODERMIC SINGLE LUMEN NEEDLE Back to Search Results
Model Number 305891
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Needle Stick/Puncture (2462)
Date of Event 12/30/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported while using bd eclipse¿ safety needle a needle detached from the hub.There was no report of patient impact.The following information was provided by the initial reporter: issue with connection of safety needle to syringes.Not secure enough.Incident where needle has come free and pricked a staff member (this was during a demonstration of the fault in the office environment- not needle-stick injury, near miss).
 
Additional Manufacturer Narrative
A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Additional Manufacturer Narrative
The following fields were corrected: d10: device available for eval no, d10: returned to manufacturer on: na.H6: investigation summary a device history record review was completed for provided lot number 2109009.The review did not reveal any detected abnormalities during the production process that could have contributed to the reported defect and all quality tests were found to be within specification.As samples were unavailable for return, twenty retained samples of the same lot number were obtained for evaluation by our quality engineer team.The retained samples were examined and no issues with the safety mechanism were identified, it was possible to correctly activate the safety mechanisms by following the guidelines in the instructions for use.Based on the investigation results, an exact cause related to the manufacturing process could not be identified for this reported incident.H3 other text : see h10.
 
Event or Problem Description
It was reported while using bd eclipse¿ safety needle a needle detached from the hub.There was no report of patient impact.The following information was provided by the initial reporter: issue with connection of safety needle to syringes.Not secure enough.Incident where needle has come free and pricked a staff member (this was during a demonstration of the fault in the office environment- not needle-stick injury, near miss).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ECLIPSE¿ SAFETY NEEDLE
Common Device Name
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
BECTON DICKINSON, S.A.
cr mequinenza
s/n
fraga, huesca 22520
SP  22520
Manufacturer (Section G)
BECTON DICKINSON, S.A.
cr mequinenza
s/n
fraga, huesca 22520
SP   22520
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13342466
Report Number3002682307-2022-00013
Device Sequence Number4164524
Product Code FMI
UDI-Device Identifier50382903058919
UDI-Public50382903058919
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K100209
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date08/31/2026
Device Model Number305891
Device Catalogue Number305891
Device Lot Number2109009
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/05/2022
Supplement Date Received by Manufacturer02/08/2022
Initial Report FDA Received Date01/25/2022
Supplement Report FDA Received Date02/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/21/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-