DATEX-OHMEDA, INC. CARESTATION 650; ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Gas/Air Leak (2946)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/28/2021
Event Type  malfunction  
Event Description
The hospital reported a leak resulting in inaccurate tidal volume.There was no patient involvement.
 
Manufacturer Narrative
A ge healthcare service representative performed a checkout of the system and confirmed the reported issue.The breathing circuit was replaced to resolve the issue.No report of patient involvement.The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.Unique identifier: (b)(4).Legal manufacturer: (b)(6).
 
Manufacturer Narrative
Additional information was received stating that the leak rate was less than 4.5l/min.A leak condition will be noted in the preop check of the equipment, as contained in the user manual.The leak may be able to be compensated for or made up with fresh gas flow.There was not a reportable malfunction.H3 other text: additional information was received stating that the leak rate was less than 4.5l/min.A leak condition will be noted in the preop check of the equipment, as contained in the user manual.The leak may be able to be compensated for or made up with fresh gas flow.There was not a reportable malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARESTATION 650
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer (Section G)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer Contact
anthony amenson
MDR Report Key13342813
MDR Text Key289020428
Report Number2112667-2022-00157
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeSA
PMA/PMN Number
K151570
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 02/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/25/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/25/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/05/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A