Model Number Y604H
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/02/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: visual analysis of the returned sample revealed that one sample, of product code y604, was received for evaluation.After investigation of the sample, short insertion was observed in the strand.The visual inspection concluded that the combination of the needle/suture was found detached since the insertion end of the suture did not meet the requirement.The defect is caused because the suture was not properly inserted inside of the needle barrel hole resulting in a pull-out defect.Based on the information currently available, the pull-out was identified during the investigation of the sample received.This product issue will be addressed through quality system.A manufacturing record evaluation was performed for the finished device batch, and no non-conformances were identified.
 
Event or Problem Description
It was reported that a patient underwent an endoscopic procedure on (b)(6) 2021 and suture was used.Before use on the patient, it was reported that the problem of pulling off suture needle had happened in the newly dispensed suture when it was opened to prepare for urological endoscopic surgery.Another like device was used to complete the surgery.Upon the evaluation of the returned sample, the visual inspection concluded that the combination of the needle/suture was found detached since the insertion end of the suture did not meet the requirement.There were no patient consequences reported.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCRYL VIO 27IN 3-0 S/A UR-6
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
avenida de las torres 7125
col salvacar
cuidad juarez
MX  
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
MDR Report Key13343651
Report Number2210968-2022-00638
Device Sequence Number11784740
Product Code GAN
UDI-Device Identifier10705031059474
UDI-Public10705031059474
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K960653
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2026
Device Model NumberY604H
Device Catalogue NumberY604H
Device Lot NumberRCMLKH
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/03/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/06/2022
Initial Report FDA Received Date01/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-