SMITH & NEPHEW, INC. LGN HK FEM ASSEMBLY SZ 7 LT; PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 71421377
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Bacterial Infection (1735)
Date of Event 10/18/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that after a revision surgery for a periprosthetic spiral fracture of the femur was performed in (b)(6) 2021 (covered under (b)(4)), a strain of e.Coli and s.Haemolyticus was detected in the patient in (b)(6) 2021.The issue required antibiotic therapy.The outcome of the patient is unknown.
 
Additional Manufacturer Narrative
D1,d2,d4,g4,h3,h6: the device was not returned for evaluation.The pictures provided were reviewed and could not confirm the infection.The clinical/medical investigation concluded that, the mechanism of contamination which led to the post-op bacterial infection could not be definitively concluded based on the documentation provided.The patient impact beyond the infection and prescribed long-term antibiotic therapy based on sensitivities could not be determine.The current patient status is unknown, although a ¿limited flexion of knee joint¿ with ¿an anesthesia mobilization¿ was documented (b)(6) 2021 just prior to (b)(6) 2021 discharge.No further medical assessment can be rendered at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the instructions for use documents for knee systems revealed that acute post-surgical wound infection and late deep wound sepsis has been identified as a possible adverse effect.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A review of the sterilization records revealed the batch was sterilized within normal parameters.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but are not limited to contamination, patient reaction, and post-operative healing issue.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN HK FEM ASSEMBLY SZ 7 LT
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13343657
Report Number1020279-2022-00309
Device Sequence Number17166405
Product Code KRO
UDI-Device Identifier00885556414880
UDI-Public00885556414880
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K081111
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source Other,Foreign,Consumer,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/25/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71421377
Device Catalogue Number71421377
Device Lot Number20LM12200
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/06/2022
Supplement Date Received by Manufacturer05/11/2022
Initial Report FDA Received Date01/25/2022
Supplement Report FDA Received Date05/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age77 YR
Patient SexMale
-
-