SMITH & NEPHEW, INC. LGN HK FEM ASSEMBLY SZ 7 LT; PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 71421377
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Anemia (1706)
Date of Event 08/04/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that after the patient underwent a 2-stage revision surgery to address a deep infection, the patient suffered from post-operative bleeding anemia.The adverse event was treated with transfusion of 2 erythrocyte concentrates with no further complications.
 
Additional Manufacturer Narrative
Internal reference number: (b)(4).
 
Additional Manufacturer Narrative
H3, h6: the device was not available for evaluation.The pictures provided were reviewed and could not confirm the stated failure mode.The clinical/medical investigation concluded that, the root cause of the first revision (2-stage) was due to a deep infection with s.Epidermidis and related symptoms; however, the mechanism of contamination which led to the deep infection could not be definitively concluded.The post-op ¿bleeding anemia¿ and subsequent transfusion was most likely due to the surgical procedure itself and does not reflect a component mal-performance.The assessed patient impact was the pain, swelling, infection, synovitis, aspirate, 2-stage revision, antibiotic therapy, blood loss anemia, and prbc transfusion.Further patient impact could not be assessed.No further medical assessment can be rendered at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the risk management file revealed this failure mode was previously identified.A review of the ifu document for knee systems provides complete guidelines of indications, contraindications, warnings and precautions and possible adverse effects that may occur preoperative, during surgery or post operative.The anticipated risk level is still adequate.A review of the sterilization records revealed the batch was sterilized within normal parameters.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes for this event could include but not limited to surgical procedure, surgical complication or patient medical history.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN HK FEM ASSEMBLY SZ 7 LT
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13343730
Report Number1020279-2022-00310
Device Sequence Number3919307
Product Code KRO
UDI-Device Identifier00885556414880
UDI-Public00885556414880
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K081111
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71421377
Device Catalogue Number71421377
Device Lot Number20GM14423
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/06/2022
Supplement Date Received by Manufacturer03/02/2022
04/29/2022
Initial Report FDA Received Date01/25/2022
Supplement Report FDA Received Date03/02/2022
05/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/13/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
71420578 / GNS II RESURF PAT 35MM; 71421385 / LEGION HK 11MM BOLT - SLEEVE; 71422167 / LGN HK DIS FEM WDG SZ7 10MM; 71423352 / LEGION HK GD MOTION ISRT 11MM SZ 4-5 LT; 71424051 / LGN PRESSFIT STEM 18MM X 160MM; 71424055 / LGN PRESSFIT STEM 22MM X 160MM
Outcome Attributed to Adverse Event Other;
Patient Age77 YR
Patient SexMale
-
-