| Model Number |
SOFTECHD+19.5 |
| Medical Device Problem Code |
Device Handling Problem (3265)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
12/03/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Lenstec received an email stating " the posterior loop of the lens was found broken when doing the surgery.The broken lens was explanted after enlarged the incision, and another softec hd was implanted.".
|
| |
|
Additional Manufacturer Narrative
|
|
A full audit of all batch documentation relating to the production of the device was performed.The audit concluded that all procedures in manufacturing and packaging of the device had been conducted correctly.Batch reconciliation was 100.0%.The visual inspection performed on the device determined that the damage to the lens is consistent with the lens being caught and pulled in the injector.The injector used was not part of lenstec's validated device delivery systems for use listed in the ifu.Lenstec cannot comment of the compatability of the injector that was used with the this lens.
|
| |
|