Model Number ROB10024 |
Device Problem
Application Program Problem: Parameter Calculation Error (1449)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/28/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that, during distal bur of femur in exposure mode during a cori assisted tka surgery, the green was not able to be taken off.The procedure was completed, with a non-significant delay, using a saw.Patient was not harmed beyond the reported problem.
|
|
Manufacturer Narrative
|
Internal complaint reference: case-(b)(4).
|
|
Manufacturer Narrative
|
Section h10: additional information received by the manufacturer has identified that this event should be re-evaluated for mdr reporting.The new information received, confirmed that the event has already been previously reported under the report reference number 3010266064-2022-00028 (our reference number (b)(4)) and that the device did not malfunction, but the cause of the issue was due to a faulty robotic drill already reported under the report reference number 3010266064-2022-00212 (our reference number (b)(4)).Therefore, it was determined that this case does not meet the threshold for reporting and will be closed as a duplicate.If further details are provided confirming the occurrence of a reportable event, our files will be updated accordingly and a further report submitted outlining both the event details and our investigations performed.Internal complaint reference number: (b)(4).
|
|