BLUE BELT TECHNOLOGIES REAL INTELLIGENCE CORI; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number ROB10024
Device Problem Application Program Problem: Parameter Calculation Error (1449)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/28/2021
Event Type  malfunction  
Event Description
It was reported that, during distal bur of femur in exposure mode during a cori assisted tka surgery, the green was not able to be taken off.The procedure was completed, with a non-significant delay, using a saw.Patient was not harmed beyond the reported problem.
 
Manufacturer Narrative
Internal complaint reference: case-(b)(4).
 
Manufacturer Narrative
Section h10: additional information received by the manufacturer has identified that this event should be re-evaluated for mdr reporting.The new information received, confirmed that the event has already been previously reported under the report reference number 3010266064-2022-00028 (our reference number (b)(4)) and that the device did not malfunction, but the cause of the issue was due to a faulty robotic drill already reported under the report reference number 3010266064-2022-00212 (our reference number (b)(4)).Therefore, it was determined that this case does not meet the threshold for reporting and will be closed as a duplicate.If further details are provided confirming the occurrence of a reportable event, our files will be updated accordingly and a further report submitted outlining both the event details and our investigations performed.Internal complaint reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAL INTELLIGENCE CORI
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13343794
MDR Text Key284375447
Report Number3010266064-2022-00051
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556757420
UDI-Public00885556757420
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K193120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROB10024
Device Catalogue NumberROB10024
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/28/2021
Initial Date FDA Received01/25/2022
Supplement Dates Manufacturer Received05/04/2022
Supplement Dates FDA Received05/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-