• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MERIT MEDICAL SINGAPORE PTE LTD. SAFEDRAW¿; SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERIT MEDICAL SINGAPORE PTE LTD. SAFEDRAW¿; SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE Back to Search Results
Model Number 682000
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/12/2022
Event Type  malfunction  
Manufacturer Narrative
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
 
Event Description
The account alleges that saline leaked out from the red stopcock adjacent to the syringe during priming before the procedure.No patient involvement.
 
Manufacturer Narrative
The suspect device has been returned for evaluation.The complaint is confirmed.No definitive root cause could be determined however, it is likely that disinfectants were applied to the device that could compromise the material.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed and no exception documents were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFEDRAW¿
Type of Device
SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE
Manufacturer (Section D)
MERIT MEDICAL SINGAPORE PTE LTD.
198 yishun ave 7
singapore, sgp 76892 6
SN  768926
Manufacturer (Section G)
MERIT MEDICAL SINGAPORE PTE LTD.
198 yishun ave 7
singapore, sgp 76892 6
SN   768926
Manufacturer Contact
bryson heaton
1600 merit parkway
south jordan, UT 84095
8012084662
MDR Report Key13344195
MDR Text Key284987715
Report Number8020616-2022-00005
Device Sequence Number1
Product Code DXN
UDI-Device Identifier00886333620005
UDI-Public00886333620005
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K935782
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/25/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/23/2023
Device Model Number682000
Device Catalogue Number687552/JPB
Device Lot NumberC2078733
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/24/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/08/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/09/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-