• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THORATEC CORPORATION HEARTMATE 11 VOLT LI-ION BACKUP BATTERY; VENTRICULAR (ASSISST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THORATEC CORPORATION HEARTMATE 11 VOLT LI-ION BACKUP BATTERY; VENTRICULAR (ASSISST) BYPASS Back to Search Results
Model Number 106128
Medical Device Problem Code Electrical Power Problem (2925)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/09/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the patient was seen in the clinic for a routine backup battery exchange due to expiration.The new backup battery out of box triggered a backup battery fault alarm.This happened with 2 additional out of box backup batteries.The fourth backup battery installed worked without issue.Additional information reported that 2 of the batteries had bent pins out of box.Related manufacturer reference number: 2916596-2022-00487 & 2916596-2022-00489.
 
Additional Manufacturer Narrative
Manufacturer's investigation conclusion: the reported event of bent pins in the backup battery was confirmed via evaluation of the returned backup battery.Visual inspection of the returned 11v backup battery (serial number: (b)(4) ) revealed that pin 11 was bent.The bent pin was straightened for testing.After straightening the bent pin, the backup battery was functionally tested.The battery was installed in a test system controller and successfully operated in a mock circulatory loop without any issues or atypical alarms produced.The controller was disconnected from external power and the battery was found to operate as intended during analysis.The battery was installed in a test system controller and successfully operated in a mock circulatory loop without any issues or atypical alarms produced.The controller was then disconnected from external power and the battery supported the pump without any issues.The root cause of the reported event could not be conclusively determined through this analysis.Heartmate 3 instructions for use section 5 ¿ ¿surgical procedures¿, explains how to install the backup battery in the system controller.Additionally, section 2 ¿ ¿system operations¿, explains how to properly replace the backup battery.Battery inspection data was reviewed for the 11v backup battery, serial number (b)(4), revealing that the battery passed all inspection testing.Customer order was shipped on (b)(6) 2021 the heartmate 3 patient handbook and the heartmate 3 instructions for use caution the user to call their hospital contacts if they think, for any reason, any portion of their equipment is not functioning as usual, is broken, or they are uncomfortable with the operation of the equipment.No further information was provided.The manufacturer is closing the file on this event.
 
Additional Manufacturer Narrative
Manufacturer's investigation conclusion: the reported event of bent pins in the backup battery was confirmed via evaluation of the returned backup battery.Visual inspection of the returned 11v backup battery (serial number: (b)(4) ) revealed that pin 11 was bent.The bent pin was straightened for testing.After straightening the bent pin, the backup battery was functionally tested.The battery was installed in a test system controller and successfully operated in a mock circulatory loop without any issues or atypical alarms produced.The controller was disconnected from external power and the battery was found to operate as intended during analysis.The battery was installed in a test system controller and successfully operated in a mock circulatory loop without any issues or atypical alarms produced.The controller was then disconnected from external power and the battery supported the pump without any issues.The root cause of the reported event could not be conclusively determined through this analysis.Heartmate 3 instructions for use section 5 ¿ ¿surgical procedures¿, explains how to install the backup battery in the system controller.Additionally, section 2 ¿ ¿system operations¿, explains how to properly replace the backup battery.Battery inspection data was reviewed for the 11v backup battery, serial number (b)(4), revealing that the battery passed all inspection testing.Customer order was shipped on (b)(6) 2021 the heartmate 3 patient handbook and the heartmate 3 instructions for use caution the user to call their hospital contacts if they think, for any reason, any portion of their equipment is not functioning as usual, is broken, or they are uncomfortable with the operation of the equipment.No further information was provided.The manufacturer is closing the file on this event.
 
Event or Problem Description
It was reported that the patient was seen in the clinic for a routine backup battery exchange due to expiration.The new backup battery out of box triggered a backup battery fault alarm.This happened with 2 additional out of box backup batteries.The fourth backup battery installed worked without issue.Additional information reported that 2 of the batteries had bent pins out of box.Related manufacturer reference number: 2916596-2022-00487 & 2916596-2022-00489.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTMATE 11 VOLT LI-ION BACKUP BATTERY
Common Device Name
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer (Section G)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer Contact
bob fryc
6035 stoneridge drive
pleasanton, CA 94588
7818528204
MDR Report Key13344637
Report Number2916596-2022-00486
Device Sequence Number17291878
Product Code DSQ
UDI-Device Identifier00813024011255
UDI-Public00813024011255
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/06/2023
Device Model Number106128
Device Catalogue Number106128
Device Lot Number8143509
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/19/2022
Initial Report FDA Received Date01/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/06/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age74 YR
Patient SexMale
Patient Weight86 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-