| Model Number |
196 |
| Medical Device Problem Code |
Suction Failure (4039)
|
| Health Effect - Clinical Codes |
Low Oxygen Saturation (2477); Unspecified Respiratory Problem (4464)
|
| Date of Event |
01/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during a tracheal aspiration procedure the closed suction catheter (csc) would not suction secretions.A 6french open suction catheter along with normal saline was used to clear the secretions from the patient's airway.The patient's vital signs decompensated and the patient's fraction of inspired oxygen (fio2) setting had to be increased for a short while; there was no report of injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The sample is reported to be available, but has not yet been received by the manufacturer.A review of the device history record is not possible as no lot number was provided.All information reasonably known as of 25-jan-2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The product involved in the report has been returned and is being processed for evaluation.A review of the device history record is in-progress.All information reasonably known as of 11 feb 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|
| |
|
Event or Problem Description
|
|
It was reported via fda medwatch / fda user facility report # mw (b)(4) the following information: tracheal aspirate being obtained by proper procedure and upon attempt to suction and clear the ballard closed suction catheter (with normal saline) it was noticed that suction (verified to be on and functioning) was not reaching the patient's airway through the catheter.Suction tubing, catheter and mucous collection specimen container were troubleshooted for leaks.Attempted to suction patient's airway again without success.Again, noted by feel and hearing, the end of the suction tubing has suction present, but the suction was not actually reaching the catheter tip while depressing the thumb button.Ballard was removed and a 6fr open suction was used to clear both the secretions and normal saline that has been inadvertently instilled into the patient's airway.Patient vitals decompensated-required fio2 for a short period.A new ballard was used to obtain an aspirate and continued clearance of the airway.Additional information received 18-jan-2022 stated the "ballard malfunction resulted in patient decompensation with need to remove ballard and suction airway manually.
|
| |
|
Additional Manufacturer Narrative
|
|
The actual sample from the reported event was returned for evaluation.Visual examination of the device did not reveal any breaks, damage or blockages.Testing of the sample revealed that the thumb valve was functioning as designed and suction was present.The reported event could not be confirmed as reported; therefore, the root cause cannot be confirmed.The device history record for lot 30142957 was reviewed and the product was produced according to product specifications.All information reasonably known as of 09 mar 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|
| |
|