• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS, S.A. DE C.V. ALARIS SMARTSITE NEEDLE-FREE VALVE; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS, S.A. DE C.V. ALARIS SMARTSITE NEEDLE-FREE VALVE; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number 2000E CHINA
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/06/2022
Event Type  malfunction  
Manufacturer Narrative
A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Event Description
It was reported while using alaris smartsite needle-free valve foreign matter was found in the fluid path.The following information was provided by the initial reporter: after the head nurse opened the outer package, there were oily water droplets in the cross section of the joint.
 
Event Description
It was reported while using alaris smartsite needle-free valve foreign matter was found in the fluid path.The following information was provided by the initial reporter: after the head nurse opened the outer package, there were oily water droplets in the cross section of the joint.
 
Manufacturer Narrative
H.6.Investigation: a 2000e china product was not available for investigation; however the customer confirmed that the complaint sample was from lot 19065675.Further information provided by the customer indicates that the oily droplets were observed on the surface of the piston of the smartsite component.A review of the production records for lot 19065675 did not identify any in-process testing failures or quality deviations which may have resulted in a report of this nature.The root cause of the customer¿s experience could not be determined as the sample was not available for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS SMARTSITE NEEDLE-FREE VALVE
Type of Device
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13347976
MDR Text Key285357597
Report Number9616066-2022-00025
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K960280
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/25/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/08/2022
Device Catalogue Number2000E CHINA
Device Lot Number19065675
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/17/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/06/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-