| Model Number |
CD2311-36 |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
01/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This report is to advise of an event observed during analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
No complaint was received with the return of the device.Failure event observed during analysis.Final analysis found evidence and confirmation of premature battery depletion.A longevity calculation was performed and found the battery depletion was premature based on the device usage.Electrical testing revealed high current drain from the hybrid which was found to be the root cause of the premature battery depletion.
|
| |
|