Model Number 48280232
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/12/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device has been discarded.
 
Event or Problem Description
It was reported that the tip of an es2 k-wire snapped off during screw insertion.The broken piece was removed and surgery was successfully completed with a 5 minute delay.No adverse consequences or medical intervention were reported.
 
Additional Manufacturer Narrative
Visual, dimensional, material and functional analysis could not be performed as the device was not returned.Device history records could not be reviewed as a valid lot number was not provided and could not be obtained.Complaint history records were reviewed for this catalog number, no similar complaints were identified.The device stg was reviewed and the following was identified: k-wire management as the awl or tap advances into the pedicle, the proximal end will move relative to the markings on the k-wire.If this does not occur during insertion, the procedure should be stopped and imaging should be used to verify the position of the k-wire in relation to the awl or tap.Note: the k-wire has 10mm increment markings to be used for reference to determine the depth of the instruments in the pedicle.Note: cantilevering the awl and taps while in the pedicle may damage the k-wire.Due to no product return or imaging of failure mode a root cause could not be determined conclusively.Cantilevering awl and taps can damage the k-wire and subsequently effect screw insertion.
 
Event or Problem Description
It was reported that the tip of an es2 k-wire snapped off during screw insertion.The broken piece was removed and surgery was successfully completed with a 5 minute delay.No adverse consequences or medical intervention were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ES2 LE NITINIOL KWIRE BLUNT
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
STRYKER SPINE-US
2 pearl court
allendale NJ 07401
Manufacturer (Section G)
STRYKER SPINE-FRANCE
zone industrielle de marticot
cestas 33610
FR   33610
Manufacturer Contact
rita karan
2 pearl court
allendale, NJ 07401
2017608000
MDR Report Key13349158
Report Number0009617544-2022-00006
Device Sequence Number1675471
Product Code LXH
UDI-Device Identifier07613327096200
UDI-Public07613327096200
Combination Product (Y/N)N
Initial Reporter CountryUK
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number48280232
Device Catalogue Number48280232
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/12/2022
Supplement Date Received by Manufacturer04/12/2022
Initial Report FDA Received Date01/25/2022
Supplement Report FDA Received Date05/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-